part # MP-A
(Allied Healthcare® / Chemetron®)
step #1
step #2
screw Medi-Purge™ into outlet / inlet
NF nitrogen will now flow out of the Medi-Purge™. An oxygen
analyzer (per NFPA-99) can be placed in the 1/4" hole, ensuring
all air has been expelled
step #3
if desired, an NF
nitrogen hose can be screwed
onto the 1/4" NPT connection allowing NF nitrogen
to flow into the outlet / inlet